SpectroQuest 4802 UV/Vis Spectrophotometers Parts

 

 

 

UNICO SQ-4802 is a Double beam design. It is a stand-alone model with 1.8 nm fixed bandwidth and has all the features that SpectroQuest line offers in a stand-alone unit. The two detectors are measuring sample and reference respectively and simultaneously for optimizing measurement accuracy. It provides excellent performance for measurements in the range of 190nm to 1100nm. It is suitable for pharmaceutical, biochemical and clinical lab applications as well as routine applications such as quantitative analyses, kinetics, spectrum scanning, multiple components and DNA/Protein. Optional PC download software and PC Windows® application software make this instrument versatile

 

SPECIFICATIONS

SpectroQuest 4802

Optical Specifications

 

Wavelength Range

190 - 1100 nm

Spectral Bandwidth

1.8 nm

Monochromator

Double beam Littrow type, grating 1200 lines/mm

Wavelength Accuracy

±0.3 nm

Wavelength Repeatability

±0.2 nm

Wavelength Calibration

automatic at start

Light Source

Tungsten Halogen/Deuterium  Lamp

Detector

Solid Silicon Photodiode

Photometric Readout

1/4 VGA 320X240 pixels backlit LCD

Photometric Range

0 - 200%T
-0.3 - 3.0A
0 - 9999C

Photometric Accuracy

±0.3%T

Zero Drift

< 0.002A per hour after warm-up

Baseline Flatness

±0.004A

Stability

0.001A/h at 500 nm

Stray Light

< 0.15%T

Scanning Speed

Hi, Med., Low. Max. 1000 nm/min

Physical Characteristics

 

Standard Sample Holder

Single Cell Holder

Analogue/Digital Output

Digital, RS232 Serial, Centronics Parallel

Dimensions

620(L) x 400(W) x 280(H) mm

Weight

24 kg / 53lbs

Shipping Dimensions

720(L) x 525(W) x 370(H) mm

Shipping Weight

28 kg / 62 lbs

Power Requirements

115V/60Hz or 230V/50Hz, switchable

Operating Temperature

Room temperature: 22 ±2°C

Humidity Range

  less than 85%

Safety Certificate

CE